Investigation of the effect of collimation on CT Dose distribution using EBT film and PMMA phantom

Authors

10.22052/3.3.43

Abstract

X-ray computed tomography is an imaging method which makes tomographic images with high contrast and resolution. However, CT-scans deliver much more radiation dose compared with other imaging techniques. The aim of this study was to evaluate two-dimensional dose distribution in a PMMA phantom as well as investigationof the effect ofcollimation factor on the dose distribution measuredusing GAF-Chromic EBT films. In this study the GAF-Chromic EBT films were cut to the dimensions of 6Í5 cm2 and exposured with different collimations (1, 3, 5, 7, 10 mm), while the other imaging parameters were fixed. After48 h post irradiation the films were reading using Power look 2100XL scanner model at the red channel. Then, the films were analyzed using MATLAB software.The optical density values ​​of each film wereatributed to the related absorbed dose. Image analysis showed that the dose distribution is non-uniform in CT-scan.Results showed that the CTDIvol measured by EBT filmswas more than that of shown on theoperator console. These differences in collimations of 1 mm and 10 mm were 13.5%1.25% and 0.7%1.25% respectively. It was also resulted that the CT dose was decresed withincrease in collimation. In addition, it was observed that the minimum and maximum dose were delivered by using  collimation factor of 10 mm and 1 mm respectively.

Keywords


[1] R.L.Dixon,K.E. Ekstrand. A Film Dosimetry System for Use in Computed Tomography. Radiology. 127 (1978) 255-258. [2] T.B.Shope, R.M.Gagne, G.C.Johnson. A method for describing the doses delivered by transmission x‐ray computed tomography. Med phys.8)1981(488-495. [3] E. Hall, D. Brenner. Cancer risks from diagnostic radiology. Br J Radiol. 81 (2014)362-378. [4] National Radiological Protection Board (NRPD), Doses from computed tomography (CT) examinations in the UK-2003 review, Report W67, P.Shrimpton, M.Hillier, M.Lewis, M.Dunn, UK(2005). [5] M.S.Pearce, J.A.Salotti, M.P.Little, K.McHugh, C.Lee, K.P.Kim, et al. Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. The Lancet. 380(2012)499-505. [6] U.S. Food and Drug Administration (FDA). Performance standard for diagnostic x-ray systems and their major components. Final rule. Federal register, Report 70,USA (2005). [7] C.H.McCollough, S.Leng, L.Yu, Cody, D.D. J.M.Boone, M.F.McNitt-Gray. CT dose index and patient dose: they are not the same thing. Radiology. 259 (2011) 311-316. [8] S.Kim, T.T.Yoshizumi, G.Toncheva, S.Yoo, F.F.Yin. Comparison of radiation doses between cone beam CT and multi detector CT: TLD measurements. Radiat Prot Dosimetry. 132 (2008) 339-345. [9] S. Kim, T. Yoshizumi, G. Toncheva, S. Yoo, F.F. Yin, D. Frush. Estimation of computed tomography dose index in cone beam computed tomography: MOSFET measurements and Monte Carlo simulations. Health phys. 98 (2010) 683-691. [10] F. Khan The physics of radiation therapy.USA (Philadephia), Lippincott Williams & Wilkins, (2003). [11] M. Saylor, T. Tamargo, W. McLaughlin, H. Khan, D. Lewis, R. Schenfele. A thin film recording medium for use in food irradiation. Int J Radiat Appl Instrum C Radiat Phys Chem. 31 (1988)529-536. [12] P.J. Muench, A.S. Meigooni, R. Nath, W.L. McLaughlin. Photon energy dependence of the sensitivity of radiochromic film and comparison with silver halide film and LiF TLDs used for brachytherapy dosimetry. Med phys. 18(1991)769-775. [13] S.C. Uniyal, U.C. Naithani, S.D. Sharma, A.K. Srivastava. Radiochromic film dosimetry of rectal inhomogeneity and applicator attenuation in high dose rate brachytherapy of uterine cervix. J Appl Clin Med Phys. 13(2012)66-75. [14] S.A. Dini,R.A. Koona,, J.R. Ashburn, A.S. Meigooni. Dosimetric evaluation of GAFCHROMIC® XR type T and XR type R films. J Appl Clin Med Phys. 6(2005) 114-134. [15] W. Bich, M.G. Cox, P.M. Harris. Evolution of the\'Guide to the Expression of Uncertainty in Measurement\'. Metrologia. 43)2006( S161-S166. [16] R. Gotanda, T. Katsuda, T.Gotanda, M.Eguchi, S.Takewa, A.Tabuchi, et al. Computed tomography phantom for radiochromic film dosimetry. Australasian Physics & Engineering Sciences in Medicine. 30 (2007) 194-199. [17] M.J. Butson, T. Cheung, K. Peter. Weak energy dependence of EBT Gafchromic film dose response in the 50kVp–10MVp X-ray range. Appl Radiat Isot. 64 (2006)60-62. [18] A. Rink, I.A. Vitkin, D.A. Jaffray. Energy dependence (75kVp to 18MV) of radiochromic films assessed using a real-time optical dosimeter. Med phys. 34(2007)458-463. [19] M. Lewis. ImPACT, Radiation dose issues in multi-slice CT scanning. technology update, ImPACT Report 3, UK,(2005). [20] M. Aweda, R.Arogundade. Patient dose reduction methods in computerized tomography procedures: A review. Int. J. Phys. Sci. 2(2007)1-9. [21] M.F. McNitt-Gray. AAPM/RSNA Physics Tutorial for Residents: Topics in CT: Radiation Dose in CT. Radiographics. 22 (2002) 1541-1553. [22] S. Brady, T. Yoshizumi, G. Toncheva, D. Frush. Implementation of radiochromic film dosimetry protocol for volumetric dose assessments to various organs during diagnostic CT procedures. Med phys. 37(2010)4782-4792. [23] J. Boivin, N. Tomic, B. Fadlallah, F.D. Blois, S. Devic. Reference dosimetry during diagnostic CT examination using XR-QA radiochromic film model. Med phys. 38(2011)5119-5129.