Re-estimation of absorbed gamma dose using PTTL in α-Al2O3:C dosimeter

Authors

10.22052/2.1.33

Abstract

Phototransferred thermoluminescence (PTTL) was used to reeestimate the absorbed gamma dose in α-Al2O3:C dosimeter. The PTTL response of this phosphor for different UVB irradiation times was investigated. The results show a linear PTTL response for gamma absorbed doses from 0.1 to 14 mGy. Because of high PTTL sensitivity in this phosphor, the low limit of re-estimated gamma dose using PTTL method is very low compared to other phosphors.

Keywords


[1] Akselrod. M.S, Kortov. V.S, Gorelova. E.A. Preparation and properties of -Al2O3:C. Radiat.Prot.Dosim. 1993, 47, 159-164. [2] Colyott. L.E, Mckeever. S.W.S, Akselrod. M.S..An integerating UVB dosemeter in -Al2O3:C. Radiat. Prot. Dosim. 1999, 85,309-312. [3] Akselrod. M.S, Kortov. V.S, Kravetsky. D.J, Gotlib.V.I. Highly sensitive thermoluminescent anion-defective -Al2O3:C single crystal detectors. Radiat. Prot. Dosim. 1990, 32, 15-20. [4] Walker. F.D, Colyott. L.E, Agersnap Larsen. N, Mckeever. S.W.S. The Wavelength dependence of light-induced fading of thermoluminescence from -Al2O3:C . Radiat. Meas. 1996, 26,711-718. [5] Mckeever. S.W.S, Akselrod. M.S, Colyott. L.E, Agersnap Larsen. N, Polf. J.C, Whitley. V. .Characterization of -Al2O3:C for use in thermally and optically stimulated luminescence dosimetry. Radiat. Prot. Dosim. 1999, 84,163-168. [6] Muniz. J.L, Correcher. V, Delgado. A. PTTL dose re-estimation applied to quality control in TLD-100 based personal dosimetry. Radiat. Prot. Dosim. 1999, 85, 63-66. [7] Akselrod. M.S. and Gorelova. E.A. Deep traps in highly sensitive -Al2O3:C TLD crystals. Nucl. Tracks. Radiat. Meas. 1993, 21, 143-146. [8] Ostar. L, Weiss. D, Kristianpoller. N. A study of photostimulated thermoluminescence in C-doped -Al2O3:C crystals. J. Phys. 1994, 27, 1732-1736. [9] Colyott. L.E, Akselrod. M.S, Mckeever S.W.S. Phototransferred thermoluminescence in -Al2O3:C. Radiat. Prot. Dosim. 1996, 65, 263-266.