<ArticleTitle>Estimation of dosimetric parameters of I-125 brachytherapy source model 6711 using GATE8.1 code</ArticleTitle>
<VernacularTitle>تخمین پارامترهای دزیمتری کپسول براکی‌تراپی I-125 مدل6711 با استفاده از کد GATE 8.1</VernacularTitle>
			<FirstPage>1</FirstPage>
			<LastPage>8</LastPage>
			<ELocationID EIdType="pii">112285</ELocationID>
			
<ELocationID EIdType="doi">10.22052/7.3.1</ELocationID>
			
			<Language>FA</Language>
<AuthorList>
<Author>
					<FirstName>زینب</FirstName>
					<LastName>فردی</LastName>
<Affiliation>دانشگاه گیلان</Affiliation>

</Author>
<Author>
					<FirstName>پیوند</FirstName>
					<LastName>طاهرپرور</LastName>
<Affiliation>دانشگاه گیلان</Affiliation>

</Author>
</AuthorList>
				<PublicationType>Journal Article</PublicationType>
			<History>
				<PubDate PubStatus="received">
					<Year>2022</Year>
					<Month>08</Month>
					<Day>21</Day>
				</PubDate>
			</History>
		<Abstract>Brachytherapy is one type of internal radiation therapy where radiation sources, which are usually encapsulated are placed as close as possible to the tumor site inside the patient&amp;#39;s body. In this technique, it is important to determine dose distribution around the brachytherapy capsule. Hereby, in this paper, dosimetric parameters of I-125 brachytherapy source model 6711 are estimated according to TG-43U1 protocol using GATE 8.1 Monte Carlo code.
In this work, we used GATE_v8.1 to calculate dosimetric parameters of the I-125 brachytherapy source model 6711. At first, validation of the GATE platform were performed by some criteria such as radial dose function, 2D anisotropy function inside liquid water according to the AAPM TG-43U1. On the other hand, since the attenuation coefficient of the sources in the water phantom is different from that of various tissues, the effects of the various tissues on the radial dose function parameter of the I-125 brachytherapy source were investigated using GATE 8.1 code.
Dosimetric parameters of simulated I-125 brachytherapy capsule show good consistency compared with the other study. The maximum average deviation was about 3.61% and 7.29% at radial dose function and anisotropy function, respectively. The relative deviation of radial dose function in the fat, muscle, breast and lung tissue compared with water phantom in  radial distance of 5cm were about 68.73%, 10.98%, 25.83% and 12.23%, respectively.
There was a good agreement between the results of this work and other study in calculation of dosimetric parameters of brachytherapy I-125 source base on the recommendations of TG-43U1 protocol. The results show that the dosimetric parameters of I-125 brachytherapy can be calculated using the GATE code and appropriate physic list in spite of low energy of radiation and high variation in dose rate with increasing distance from the center of the source. The results of the dose calculation in different phantom could be used in clinical treatment planning systems.

 </Abstract>
			<OtherAbstract Language="FA">براکی‌تراپی نوع خاصی از  پرتودرمانی است که در آن چشمه پرتوزا در داخل بدن بیمار و در نزدیکی (و یا در داخل) تومور قرار می‌گیرد. در این بین، تعیین توزیع دز پیرامون چشمه‌های مورد استفاده در براکی‌تراپی برای اهداف طراحی درمان بسیار با اهمیت است. در این تحقیق، پارامترهای دزیمتری توصیه شده توسط پروتکل TG43-U1، AAPM، نظیر ثابت آهنگ دز، تابع دز شعاعی و تابع ناهمسانگردی چشمه براکی‌تراپی I-125 Amersham Health مدل 6711 با استفاده از کد شبیه‌سازی  مونت‌کارلو GATE در فانتوم آب محاسبه و اعتبارسنجی شده است، و در ادامه، با توجه به ساختار متفاوت بافت‌های مختلف بدن نسبت به فانتوم آب، نتایج مربوط به محاسبات دزیمتری، به بافت‌های چربی، سینه، ریه، دو نوع بافت نرم و ماهیچه تعمیم داده شد. شبیه‌سازی‌ها نشان‌دهنده تفاوت محسوس دز جذبی در برخی از بافت‌ها، نسبت به فانتوم آب (که توسط پروتکل استاندارد TG43-U1، AAPM پیشنهاد می‌شود) است، به طوری که تغییرات مربوط به تابع دز شعاعی در بافت چربی در فاصله cm 5 از مرکز چشمه به 7/68% نیز خواهد رسید. نتایج به دست آمده بیانگر آن است که پارامترهای دزیمتری چشمه براکی‌تراپی 125-I را می‌توان، علی‌رغم انرژی تابشی پایین و تغییرات شدید دز با افزایش فاصله از مرکز چشمه، با استفاده از کد GATE و فهرست فیزیکی مناسب محاسبه کرده ودر راستای توسعه نرم‌افزارهای طراحی درمان براکی‌تراپی مورد استفاده قرار داد.
 </OtherAbstract>
		<ObjectList>
			<Object Type="keyword">
			<Param Name="value">براکی‌تراپی</Param>
			</Object>
			<Object Type="keyword">
			<Param Name="value">چشمه 125-I</Param>
			</Object>
			<Object Type="keyword">
			<Param Name="value">دزیمتری</Param>
			</Object>
			<Object Type="keyword">
			<Param Name="value">کدGATE</Param>
			</Object>
			<Object Type="keyword">
			<Param Name="value">پروتکل TG43-U1</Param>
			</Object>
		</ObjectList>
<ArchiveCopySource DocType="pdf">https://rsm.kashanu.ac.ir/article_112285_460e50ce2f4b35140f81dda6cb53e4d1.pdf</ArchiveCopySource>
</Article>
</ArticleSet>